Development of Low Pressure Position Sensitive Multiwire Proportional Chamber
- Received Date: 2003-04-17
- Accepted Date: 1900-01-01
- Available Online: 2003-11-05
Abstract: A low pressure two dimensional position sensitive multiwire proportional chamber (LPMWPC) has been developed for measurements of low Z charged particles from the nuclear reaction induced by radioactive ion beam. The detector has a 50mm×50mm sensitive area and consists of three grids (X-position cathode Y-position) . The position is determined through delay line readout mode. The detector has good beam transparency and is also adequate to be used in vacuum . The LPMWPC has been tested at a pressure of 800Pa of n-heptane and a bias Voltage of 720V. The test results indicate that the position resolution of the chamber is Δx = Δy = 1mm and detection efficiency is larger than 80% for the low Z particle at intermediate energies. This kind of chamber has be used to determine the direction of the incident particle.