THE INVESTIGATION OF THE HIGH DENSITY MULTIWIRE PROPORTIONAL CHAMBER AS A DETECTOR FOR NEUTRAL PARTICLES
- Received Date: 1900-01-01
- Accepted Date: 1900-01-01
- Available Online: 1987-10-05
Abstract: This paper describes a position sensitive multiwire proportional chamber with effective area of 200×200mm2. The chamber combining with multi-layer lead γ-ray converter form a high density multiwire proportional chamber γ-ray detector. The lead converter is made in auniformly arranged matrix of 0.9mm diameter holes on a 1.1mm pitch. For 511keV photons the detection efficiency is 5.6%, and the spatial resolution is 1mm. The detector has been used in the investigation of positron camera, it also offers possibility for X-ray detection.