Calibration of Energy Response of Detection Elements for Soft X-Ray Spectrometer
- Received Date: 2000-02-02
- Accepted Date: 1900-01-01
- Available Online: 2001-05-05
Abstract: The calibration of the energy response of the detection elements for soft X-ray spectrometer is reported. The calibration is carried out Beamline 3W1B of Beijing Synchrotron Radiation Facility with beam intensity 40—80mA at electron energy 2.2GeV,using vacuum chamber of reflectometer on An AXUV-100 silicon photodiode is used as the source intensity absolute monitor in the photo energy rang 150—1500eV. The energy response curves for X-ray diode (XRD) quantum efficiency,filter transmittance and grazing planar mirror reflection efficiency have been measured. Finally,the possible maximum errors of the XRD at different energies are given based on these experimental data.