Restoration of the Signals from Multi-hit Three-Layer Delay-Line Anode Position-Sensitive Detector
- Received Date: 2006-05-11
- Accepted Date: 2006-07-10
- Available Online: 2007-03-05
Abstract: Multi-hit 3-layer delay-line anode (Hexanode) has an increased ability to detect multi-hit events in a collision experiment. Coupled with a pair of micro-channel plates, it can provide position information of the particles even if the particles arrive at the same time or within small time dwell. But it suffers from some ambiguous outputs and signal losses due to timing order and triggering thresholds etc. We have developed a signal reconstruction program to correct those events. After the program correction, the dead time only exists when 2 particles arrive at the same time and the same position within a much smaller range. With the combination of Hexanode and the program, the experimental efficiencies will be greatly improved in near threshold double ionization on He collisions.