The Efficiency Mearurement of NaI(Tl) Detector for a Long Tube Volume Source of 6.13MeV Gamma-Ray
- Received Date: 2004-06-07
- Accepted Date: 1900-01-01
- Available Online: 2005-02-05
Abstract: The relation of the 16N leak rate of the nuclear reactor to the efficiency and counting rate of γ–ray detector, velocity of high pressure vapor, is deduced. a conclusion of the efficiency independent on the velocity of high pressure vapor and decay constant of 16N is drawn. experimentally the space distribution of the relative efficiencies and the absolute efficiency at a refe~rence point are measured with Puc source emiting 6.13mev gamma ray detected by nai(tl) counter. The measured efficiency is consistent with monte carlo calculation in 7%. the compton small angle scattering contribution to the convergence length and the effect of the energy difference between 7.12mev and 6.13mev gamma ray on measured 16N leak rate are discussed.