Preliminary study of depth of interaction measurement for a PET detector
- Received Date: 2011-01-24
- Accepted Date: 2011-03-02
- Available Online: 2011-11-05
Abstract: In this work we studied the feasibility of detecting the depth of interaction (DOI) with two layers of crystal arrays of LYSO and BGO scintillators coupled to a position-sensitive photomultiplier tube (PS-PMT) R8900-C12. A front-end electronics was designed, with which we got different pulse shapes for different crystals to obtain depth information. With the double integration method, we got the DOI histogram of a divided integration ratio of two crystals as the standard to determine the layer-of-interaction. The DOI accuracy, measured by scanning a 22Na slit source along the side of the module, was 98% for the LYSO layer and 95% for the BGO layer. The energy resolution at 511keV was 13.1% for LYSO and 17.1% for BGO. We obtained good crystal separation in 2D position histograms of both layers. These results could be useful in the manufacture of PET scanners with high spatial resolutions.