THE MONITOR SYSTEM OF COUNTERS OF MARK-J DETECTOR

Get Citation
ZHENG ZHI-PENG. THE MONITOR SYSTEM OF COUNTERS OF MARK-J DETECTOR[J]. Chinese Physics C, 1982, 6(2): 135-141.
ZHENG ZHI-PENG. THE MONITOR SYSTEM OF COUNTERS OF MARK-J DETECTOR[J]. Chinese Physics C, 1982, 6(2): 135-141. shu
Milestone
Received: 1981-06-08
Revised: 1900-01-01
Article Metric

Article Views(1448)
PDF Downloads(236)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

THE MONITOR SYSTEM OF COUNTERS OF MARK-J DETECTOR

  • Institute of High Energy Physics, Academia Sinica

Abstract: A xenon lamp-fibre glass apparatus as a monitor system of counters of Mark-Jdetector has been built. There are 208 channels of fibre to photomultipliers of counters.The calibration of pulse amplitude of counters and the relation of time between coun-ters has been given. Due to the reference tube system, the stability of the apparatus has been improvedand the accuracy of measurement has approached to ±1%. After long time operation, we found that the apparatus are stable and reliabledesirable.