-
Located at China Jin-Ping Underground Laboratory (CJPL) [11], the PandaX-II experiment has been taking physics data since 2016. The detector is a dual-phase xenon TPC [12], as shown in Fig. 1. Detailed description of the detector can be seen in [8]. The sensitive volume is defined by a cylinder of polytetrafluoroethylene (PTFE) reflectors of inner diameter 646 mm and two sets of photomultiplier tube (PMT) arrays at the top and bottom. The drift/extraction/amplification field in the TPC is exerted by the cathode, gate, and anode meshes, and is constrained by a set of copper shaping rings outside the reflector barrel. Particle going through the liquid generates scintillation photons and ionized electrons, the latter of which drift to the gas phase and generate electroluminescence photons. Light signals are used to reconstruct energy and position information of a particular event. Vetoing volume resides outside of the PTFE reflectors and is used to reject environment and detector background events originating from outside.
Figure 1. (color online) Schematic drawing of the PandaX-II TPC. Key components such as top and bottom PMT arrays, PTFE reflectors, and electrodes are highlighted. The figure is adapted from [8]
Multiple layers of passive shielding outside the TPC include copper, inner polyethylene, lead, and outer polyethylene (see Fig. 2). Bricks or panels of those materials are concatenated into octagonal shape. The gap between the innermost layer and the TPC vessel is flushed by nitrogen gas in order to suppress radon concentration inside. Ambient radiation such as gamma and neutron background, is attenuated effectively with the help of passive shielding.
Figure 2. (color online) Schematic drawing of the layered passive shielding system around the PandaX-II Detector. Thickness of different layers are labelled. The figure is adapted from [13]
In this paper, three sets of physics data from the past 3 years are used for NLDBD analysis, as shown in Table 1. The total accumulated exposure is 403.1 days. Run numbers in the table follow the convention used in dark matter direct detection physics analysis. Run 9 and 10 are the first two sets of physics runs for PandaX-II and have been previously published [9, 10]. Run 11 is a continuation of Run 10 with no change in run conditions. Run 11 data taking was undertaken smoothly from July 2017 to August 2018, interrupted only by a series of calibration runs, as shown in Fig. 3. The figure also shows that the drift electron lifetime, which is an indication for detector performance, was stable at above 400 μs most of the time. During this run, an additional 246.4 live-days of data was accumulated. This is the first time that we present an analysis with this new data set.
data set begin end live days Run 9 Mar. 9, 2016 Jun. 30, 2016 79.6 d Run 10 Apr. 22, 2017 Jul. 16, 2017 77.1 d Run 11 Jul. 17, 2017 Aug. 16, 2018 246.4 d Table 1. PandaX-II data sets used for this NLDBD analysis.
Figure 3. (color online) Data taking history of Run 11. Other than 4 interruptions of calibration and detector performance study, we took data continuously from July 2017 to August 2018. The left Y axis is the exposure in the unit of kg·yr. The dash horizontal line indicates the total exposure at 242 kg·yr, with a corresponding fiducial mass of 219 kg. The right Y axis is the electron life time while drifting in the field cage. The electron life time is measured to be stable in the last period of Run 11. The average value is used for position uniformity correction (explained later in the text) during energy reconstruction and shown in the figure.
-
PandaX-II data analysis at MeV level follows the workflow of dark matter analysis closely [14]. However, we developed new algorithms for energy reconstruction and new event selection criteria at MeV level specifically for NLDBD analysis. In this section, we summarize the analysis workflow and highlight the key differences.
-
PandaX-II data acquisition system (DAQ) records triggered waveforms of prompt scintillation light signal (commonly called
$ S1 $ ) and electroluminescence light signal ($ S2 $ ). For each triggered signal, the readout window is 1 ms and the trigger point is placed in the middle of the signal window. Data are recorded by underground DAQ computers and immediately copied to a local computing farm aboveground at CJPL, where first level data preprocessing is performed in situ automatically.The first step of data preprocessing chain process waveform of individual PMT. After the Single Photoelectron (SPE) correction, which we will describe later, we group these waveforms into individual pulses, known as hits, using a hit finding algorithm.
The second step is to group coincident hits of PMTs to construct light signals. Waveforms from individual pulses are summed up and the key parameters are calculated for the summed pulse. We then classify these signals into
$ S1 $ ,$ S2 $ or noise using a decision tree based on the characterization of summed waveforms. A typical event may include one$ S1 $ signal and one or multiple$ S2 $ signals. Each$ S2 $ signal paired with the preceding$ S1 $ signal corresponds to one interaction vertex (usually called site) in the liquid xenon medium.Finally we reconstruct the positions of sites in each event. The Z direction follows electron drift direction and points upward while X and Y directions define the horizontal plane. The X and Y coordinates of each track are calculated from
$ S2 $ signal. Multiple methods are used in this step, including center of gravity (COG), template matching (TM), and maximum likelihood finding using photon acceptance function (PAF) [9], the last of which gives best uniformity of event distribution of calibration source, and is used as standard position reconstruction algorithm in the following analysis. The Z coordinate is calculated from electron drift velocity in the electric field and the drift time dt, defined as time difference between$ S1 $ and$ S2 $ signals. In Run 9 the drift velocity is$ 1.71\pm0.01 $ mm/μs with an electric field of 393.5 V/cm. In Run 10 and Run 11, the drift velocity dropped to$ 1.67\pm0.01 $ mm/μs due to the lowered drift field of 311.7 V/cm. The derived drift velocities are consistent with Ref. [15]. -
The detector's response at MeV level is calibrated using external 232Th source capsules. The source is deployed through the calibration loops inside the TPC outer vessel. A full absorption gamma peak from 208Tl at 2615 keV can be seen in both
$ S1 $ and$ S2 $ signals. For MeV scale signals, a large$ S2 $ happens right below the top PMT array and the number of photons collected by a PMT may be beyond the linear response region. In rare cases when signal size exceeds the dynamic range of digitizers, the waveform is also clipped. Saturation of PMTs and digitizers deteriorates energy resolution. A group of saturated pulses with both PMT and digitizer saturation is shown in Fig. 4.Figure 4. (color online) Examples of saturated waveform of PMTs. Different traces show waveforms from different top PMTs from a high energy event. The saturated PMT (black and yellow line) has asymmetric peak in its waveform, where the rising edge is steeper compared to the other ones. For PMT 10900, a typical PMT that experiences saturation problem, clipped peak is also observed as the digitizer saturates.
In our analysis, we used pulses from the bottom PMT array to reconstruct the energy of
$ S2 $ signals to avoid the saturation problem. Saturation is not observed in the bottom PMT array, since$ S1 $ signal is usually much smaller than$ S2 $ and$ S2 $ signal is far away from the bottom array. We describe the whole process below and emphasis is given on last two steps, where pulse saturation has key impact.SPE correction Hit pulses are scaled by the individual SPE values of PMTs so that the area of pulse represents the number of photoelectrons (PE). This correction compensates possible PMT gain drift during detector operation.
Position dependent correction The number of PE that each PMT collects for any specific event depends on the X, Y, and Z position of the interaction vertex. The position dependence is mapped out with internal radioactive sources such as 83Kr or activated 127mXe. 127mXe emits 164 keV gamma rays and is better suited for position dependence correction at the high energy region. We divide the sensitive volume into small voxels to map out detector response at different position. In turn PMT response for any events is corrected with the mapping.
Secondary electrons from xenon ionization are recombined with or attached to the impurities in the drifting process. The electron loss is usually characterized by electron life time in the unit of μs. The measured
$ S2 $ signal at liquid-gas interface is corrected with electron life time on a daily basis.Energy reconstruction To reconstruct the energy of an event, we follow the energy sharing formula used in dark matter analysis:
$ E = W \times \left(\frac{S1}{PDE}+\frac{S2}{ {EEE} \times {SEG}}\right), $
here W equals to 13.7 eV and is the average work function for a particle to excite one electron-ion pair or one UV photon in xenon [16].
$ S1 $ is the total detected PE number from both top and bottom PMTs, after event position correction. Since the saturated top PMT charge cannot be used, we replace$ S2 $ with bottom PMT PE number:$ S2 = S2_{ {\rm Bottom}} \times (T/B+1), $
where
$ T/B $ stands for ratio between the numbers of PE from the top and bottom PMT array. We expect that photon collection is energy independent and set$ T/B = 2.3 $ , as derived in the dark matter analysis [9]. The other three parameters, photon detection efficiency (PDE), electron extraction efficiency (EEE), single electron gain (SEG), are derived by scanning the parameter space as follows. For each combination of PDE and EEE$ \times $ SEG, a spectrum of 232Th calibration source is obtained and fitted. The combination of 7.4% and 22.75 PE/e yields the best energy resolution and is selected. The achieved best energy resolution is 4.2% at 2615 keV. The PDE and EEE$ \times $ SEG are different from those used in dark matter analysis, which implies changes of detector response in this different energy region. The new set of values is shown as the red line in Fig. 5 overlaid with the$ S1 $ vs.$ S2 $ bottom signals.Figure 5. (color online) Anti-correlation between
$S1$ signals (Top + Bottom PMTs) and$S2$ Bottom PMT signals in 232Th calibration data. The bright blob on the top-right is the 2615 keV gamma peak. The red line indicates the best anti-correlation between$S1$ and$S2$ signals, whose slope is calculated from the ratio of PDE and EEE × SEG.Residual non-linearity correction A spectrum with optimized energy resolution can be obtained with the above mentioned procedures. However, peaks could be identified slightly off the true energy of the characteristic gamma lines. Another correction is carried out to shift those peaks and correct the non-linearity of detector's energy response. With a relatively large energy resolution, gamma lines on top of continuum background may not peak in the exact gamma energy. Monte Carlo simulation is used to study such shifts. The detailed simulation process is discussed in Sec. 4. Peaks from 915 to 2615 keV are used to correct the residual non-linearity, as shown in Fig. 6. For each run, six peaks are used to fit a third-order polynomial function with zero intercept. Then energy of each event is corrected with these fit functions.
Figure 6. (color online) Peak matching and residual nonlinearity of the three runs. The X axis is the detected peak position from 915 to 2615 keV before corrections. The Y axis is the expected peak position from Monte-Carlo simulation. Three-order polynomial functions are used to fit the data to determine the correction functions for corrections.
-
The original cuts from dark matter analysis are inherited but tuned to adapt to the high energy region in this study. We first apply the single-site cut, ROI energy cut, veto cut, and fiducial volume cut to reject a vast majority of background events. The signal quality cuts, including a few pulse shape cuts,
$ S1/S2 $ band cut, top/bottom asymmetry cut, and position quality cuts, are used to further reject possible low quality signals. Here we briefly describe the cut criteria and more detailed definition can be found in [13, 14].Single-site cut We focus on single-site (SS) events from NLDBD and reject all events with multiple
$ S2 $ signals, because in most cases the two electrons emitted from an NLDBD event deposit energy in one point-like site in liquid xenon. Monte Carlo shows that such selection accepts 93.4% of all the NLDBD events, and the inefficiency is dominated by the multi-site (MS) deposition due to the bremsstrahlung. On the other hand, background events from gammas are predominately multi-site events, the cut reduces gamma background effectively. For example, in run 10, 6.9 million out of 18.4 million of triggered events are tagged as single-site events.ROI cut In the NLDBD final spectrum fit, we use events with energy in the range of 2058 to 2858 keV, i.e.
$ \pm 400 $ keV around the Q-value.Veto cuts The large majority of NLDBD events inside the field will not deposit any energy outside. On the contrary, penetrating particles such as muon may deposit energy in the veto region and is rejected with an offline coincidence analysis between veto signal and
$ S1 $ signal.Fiducial volume cut Fiducial volume for NLDBD analysis is optimized according to the distribution of high energy events. Fig. 7 shows the spatial distribution of high energy events in the ROI. The X axis shows square of event radius. The Y axis is the negative drift distance, which is equivalent to the relative vertical coordinate in space. Number of events per bin remains relatively constant along the radius direction since xenon self-shielding is not as effective for high energy events. Abnormal "stripes" of excess events are observed vertically, which is due to malfunctioning of PAF position reconstruction algorithm with saturated PMT pulses. This problem also leads to a decrease of event density at the edge of the sensitive volume. To be conservative, we force
$ r^2<80000\,\, {\rm mm}^2 $ . More background events concentrate at the top of the sensitive volume where the top PMT array, stainless steel flanges and pipes contribute significantly. Therefore, the fiducial volume cut in vertical direction is not symmetric. The vertical cut is within -233 and -533 mm, optimized by comparing fiducial mass and the number of background events under different ranges. The final fiducial region is shown in the red rectangle in Fig. 7 and the fiducial mass within is 219 kg. The fiducial volume cut is the strongest cut in the event selection, reducing large amount of background events by the self-shielding effect of liquid xenon.Figure 7. (color online) Spatial distribution of ROI event in the PandaX-II TPC. The Y axis is in the electron drift direction with the liquid and gas interface at 0 mm. Single scatter cut, quality cut, and veto cut are added. Fiducial volume used in the NLDBD analysis is marked as red rectangle.
We estimate the uncertainty of the fiducial mass with position reconstruction uncertainties. The uncertainty of position reconstruction in the radial direction is 21.5 mm and given by the mean difference between PAF and TM method for events in the ROI. The uncertainty in the vertical direction is 3 mm, mainly from the uncertainty of drift velocity measurement. Overall, we obtained an uncertainty of 32 kg for our fiducial mass.
Pulse shape cuts Two pulse shape cuts are used to reject pulses with positive overshoot. Ripple ratio cut calculates the overshoot of
$ S1 $ waveforms and rejects pulses with a ratio larger than 0.5%. Negative charge cut compares two methods of total PMT charge calculation. A discrepancy between them also indicates unwanted overshoot in corresponding waveforms. We reject events with such discrepancy larger than 1.5%.$S1/S2$ band cut Noise signals of electrode discharge have pulses similar to$ S2 $ but with narrower width. Consequently, events with small "S2" are rejected, whose log$ _{10}(S2/S1) $ are smaller than 1.5.Top/bottom asymmetry cut The cut is defined as the difference between numbers of PE collected by top and bottom PMT arrays for
$ S1 $ signals. Random coincidence events, usually with large asymmetry, are rejected.Position quality cuts This group of cuts reject events with poor position reconstruction quality. Events with unsatisfactory PAF fitting quality are rejected. We also require the difference between positions reconstructed by PAF and TM methods to be less than 40 mm.
Impacts of these successive cuts are shown in Table 2 and Fig. 8. In Table 2 we show the number of survived events and estimated signal efficiencies of the relevant cuts. Detector trigger efficiency at high energy region is estimated to be about 100% and neglected here. With a wide ROI, almost all the NLDBD events would fall in the region and the ROI cut efficiency is neglected. Veto cut efficiency is not needed since we only study single-site NLDBD events with full energy deposition. Fiducial volume cut is reflected in the final136Xe mass calculation. Efficiencies of signal quality cuts are evaluated conservatively by comparing the number of events before and after applying corresponding cut. For example, the pulse shape cut efficiency is
$ 99.3\pm0.8 $ %, averaged over the values of three runs. The uncertainty is derived from the standard deviation of cut efficiencies of three runs. By combining all the cuts, we calculate NLDBD detection efficiency to be$ 91.6 \pm 0.8 $ %. Meanwhile, the number of background events has been suppressed by three orders of magnitude.events signal efficiency (%) single scatter cut 38070871 93.4 ROI cut 771356 ≈100 veto cut 433482 − fiducial volume cut 32132 − pulse shape cut 31900 99.3 ± 0.8 $S1/S2$ pattern cut31837 99.8 ± 0.1 position quality cut 31511 99.0 ± 0.1 total − 91.6 ± 0.8 Table 2. Event selection tree and the efficiency for NLDBD signal. Signal efficiency stands for the ratio of NLDBD events that can pass through the step.
Figure 8. (color online) Energy spectrum from 0 to 3 MeV after various cuts. Impacts of different cut can be seen in the plot. Backgrounds are rejected effectively by veto and fiducial volume(FV) cuts. The signal quality cut is tuned specifically for high energy events, and yields a lower efficiency for low energy events far away from the ROI.
-
The background model of NLDBD ROI differs from the background pre-fit as described in the previous section in a few key aspects. In the ROI fit between 2058 to 2858 keV, energy resolutions are fixed to their values obtained in the pre-fit. We ignore the negligible 40K and 2NDBD contributions in this fit. Event rates from other isotopes are still to be fitted, but assigned with Gaussian priors with their sigma from the pre-fit results.
The fitted NLDBD rate
$ \Gamma_{0 \nu} $ is in unit count per year per nucleus. To add it into model, we convert the decay rate to total count C:$ C = \frac{\Gamma_{0 \nu} \times T \times N_{\rm A} \times R \times M \times \epsilon }{m} , $
where T is the detector live time,
$ N_{\rm A} $ is the Avogadro constant, R = 8.9% is the natural isotopic abundance of 136Xe, M is the xenon mass in our fiducial volume,$ \epsilon $ is the detection efficiency, and m is the molar mass of xenon.The best fit model combines data of three runs in one BAT fit. Background rate is independent in each of the runs, while the NLDBD decay rate is forced to be one parameter. In total we have
$ 3\times4+1 = 13 $ (3 runs, 4 background components) fit parameters for the model. The fitted ROI spectrum is shown in Fig. 10. Fit results of all free parameters are listed in Table 3. The fitted NLDBD rate is$ (-0.25\pm0.21) \times 10 ^{-23}\,{\rm yr}^{-1} $ , with an equivalent event rate$ -0.93\pm0.79 $ per kg·yr. The fitted event rates of 232Th and 238U are correlated with 222Rn and fluctuate in different runs. However, when forcing the Th/U event rates in the three runs to be the same, we observe effectively no change in the fitted NLDBD event rates.Figure 10. (color online) (Bottom) Energy spectrum and best-fit model for the NLDBD ROI. Three runs are fitted simultaneously with shared NLDBD rates and the spectrum is the summation of all the data. The fits are dominated by 2204 and 2615 keV gamma peaks from 238U and 232Th decay chain respectively. An NLDBD rate of
$-0.25\pm0.21 \times $ $10 ^{-23}\,{\rm yr}^{-1}$ is obtained from our best fit model. (Top) The normalized residuals of the fits.item run 9 run 10 run 11 60Co 14.5 ± 1.0 13.3 ± 1.0 11.6 ± 0.6 222Rn chain 33.4 ± 6.3 31.9 ± 5.2 22.9 ± 3.3 232Th chain 63.6 ± 2.4 55.8 ± 2.0 67.6 ± 1.3 238U chain 36.6 ± 4.2 34.8 ± 3.4 41.0 ± 2.1 NLDBD −0.93 ± 0.79 Table 3. Summary of best-fit parameters. Event rates are in the unit of counts per kg·yr.
We also estimate systematic uncertainties induced by different parameters, as listed in Table 4. As in Ref. [24], we split the uncertainty contribution into additive part (
$ \sigma_{{\rm add}} $ , which is independent of the decay rate) and scaling part ($ \sigma_{{\rm scaling}} $ , which contributes to an uncertainty proportional to the true decay rate). Uncertainties from the signal detection efficiency, fiducial mass, and MC rate (SS/(SS+MS) ratio) only have the scaling part, and are firstly added to the table.values uncertainty additive sys./ (10−23 yr−1) scaling sys. signal detection 91.6% 0.8% 0.83% fiducial mass 219 kg 32 kg 14.7% MC rate − 9% 9% MC shape − 6% 0.083 0.49% energy scale 94%-100% 1% 0.045 − energy resolution 4.2%-4.8% 0.1% 0.089 2.47% fit bias − − 0.041 0.04% Table 4. Summary of the additive and scaling systematic uncertainties of different contributions. The range of energy resolution and energy scale values used in the three runs are shown.
We carry out a set of pseudo-experiments with toy Monte-Carlo spectrum. Toy spectra are generated according to the fitted background model and mixed with varied NLDBD events. For example, for energy resolution
$ \Sigma $ , we modify its value by its uncertainty in the background model to be$ \tilde{\Sigma} $ . Hypothetical NLDBD decay rate$ \tilde{\Gamma} $ is changed from 0 to$ 3\times10^{-23} \,{\rm yr}^{-1} $ . For each fake NLDBD rate$ \tilde{\Gamma} $ we generate 1000 toy Monte-Carlo spectra according to the fitted background model but with$ \tilde{\Sigma} $ . Then the spectra are re-fitted by imposing the best fit background model with all systematic terms fixed at their best fit values (e.g.$ \Sigma $ ), yielding an NLDBD event rate which is biased from the input. The fitted NLDBD rates are plotted against the input ones and the distribution is fitted with a linear function. Interception of the fit on the Y axis represents the$ \sigma_{{\rm add}} $ and its slope's deviation from one is$ \sigma_{{\rm scaling}} $ .The energy scale uncertainties, estimated to be 1% at Q-value, also contribute to the uncertainty of the NLDBD rate. We modify the energy scale by ±1% when producing the pseudo-experiment data, and repeat the constrained fits mentioned above. For Monte Carlo shape uncertainty, we modify the height of each characteristic peak either by +6% or -6% in the simulated spectrum, and repeat the steps above. The uncertainty of the fitting process itself, known as fit bias, is derived similarly but with no change of the systematic sources.
Results of systematic uncertainty evaluation are listed in Table 4. Under the assumption that all the uncertainties are un-correlated, the total systematic uncertainty can be calculated as:
$ \sigma_{\rm{ syst }}\left(\Gamma_{0 v}\right) = \sqrt{\sum\limits_{i} \left(\sigma_{{\rm add}, i}^{2}+\sigma_{{\rm scaling}, i}^{2} \Gamma^{2}_{0 v}\right)}~. $
Our final NLDBD rate is
$ (-0.25 \pm 0.21\,({\rm stat.}) \pm $ 0.14$ \, ({\rm sys.})) \times 10^{-23}\,{\rm yr}^{-1} $ . The result is consistent with zero and we find no evidence of NLDBD in the PandaX-II data. We can calculate a physical Bayesian limit from this best-fit NLDBD rate. The posterior probability distribution and the systematic uncertainties are combined, as shown in Fig. 11. By integrating the positive part of the total likelihood curve to an area of 90% [25], we obtain a limit on NLDBD rate$ \Gamma_{0 \nu}< 0.33 \times 10^{-23} $ $ {\rm yr}^{-1} $ and NLDBD half-life$ T_{1/2}^{0\nu}> 2.1 \times 10^{23} $ yr at the 90% confidence level (C. L.). The half-life limit would be$ 2.9 \times 10^{23} $ yr if we only consider the statistical uncertainty. Effective Majorana mass can be evaluated under the assumption of light Majorana exchange in the NLDBD process. We adopt the phase space factors from [26] and the nuclear matrix elements calculated from models in [27-31]. The upper limit of effective Majorana neutrino mass$ m_{\beta \beta} $ is found to be in the range of 1.4 and 3.7 eV at the 90% C. L..Figure 11. (color online) NLDBD rate probability distribution. The red scattered dots are from BAT fit result, while the red line is their gaussian fit.
The sensitivity of
$ ^{136} $ Xe NLDBD search in PandaX-II experiment can also be calculated. We carry out another set of pseudo-experiments, whose spectra are analyzed in the same way as our experimental data. The fitted probability distribution curves are associated with the systematic table and NLDBD half-life limits at the 90% C. L. are calculated. A total of 12000 MC spectra are analyzed and the median value of derived half-life limits is chosen as the sensitivity. The resulting NLDBD half-life sensitivity is$ T_{1/2}^{0\nu}> 1.6 \times 10^{23} $ yr at the 90% C. L..
Searching for neutrino-less double beta decay of 136Xe with PandaX-II liquid xenon detector
- Received Date: 2019-06-28
- Available Online: 2019-11-01
Abstract: We report the Neutrino-less Double Beta Decay (NLDBD) search results from PandaX-II dual-phase liquid xenon time projection chamber. The total live time used in this analysis is 403.1 days from June 2016 to August 2018. With NLDBD-optimized event selection criteria, we obtain a fiducial mass of 219 kg of natural xenon. The accumulated xenon exposure is 242 kg·yr, or equivalently 22.2 kg·yr of 136Xe exposure. At the region around 136Xe decay Q-value of 2458 keV, the energy resolution of PandaX-II is 4.2%. We find no evidence of NLDBD in PandaX-II and establish a lower limit for decay half-life of 2.1